Machine technical sheet


Immagine rappresentane il macchinario
Brand: THE MAGSTIM COMPANY Ltd
Model: Magstim Rapid 2
Type: Electromedical equipment: transcranial stimulators
Power supply: Electrical (220V-380V)
Reference standard: CEI EN 60601-1
Frequency of work: 0 - 1000 Hz
Type of emission: Pulsata
Type of radiant element: Induttivo
Category according to 12198-1-12009 Standard: Non disponibile
Required protection measures:
  1. Delimit Zone 1: Exposure values above the reference levels for general public
  2. Delimit Zone 2: Exposure values above the action levels for workers
  3. The access to the Restricted areas shall be permitted only to those individuals with proven absence of contraindications to EMF exposure
  4. The access to the Restricted area is forbidden to individuals wearing pacemakers and/or implanted electronic devices

Measurements in the field (Click to view the measurements in the field)

Reference person:

AUSL Toscana Sud Est-LSP-Lab. Agenti Fisici Iole Pinto - Nicola Stacchini - Andrea Bogi ([email protected])
SIZE AND SHAPE OF THE APPLICATOR [m]
X 0.13 SHAPE

Cylindrical
Y 0.13
Z 0.01
Measuring point Reference Distance Perpendicular [cm] Distance Parallel [cm] Height from ground [cm] Frequency dominant [Hz] FIGURE MEASURED ENTITY MEASURED VALUE NOTES
P1
applicatore
50 0 0 1000 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 % indici calcolati con picco ponderato
P1
applicatore
50 0 0 1000 Als Workers UE Directive 2004/40/EC [Magnetic Field] 30 % indici calcolati con picco ponderato
P2
applicatore
80 0 0 1000 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 20 % indici calcolati con picco ponderato
P2
applicatore
80 0 0 1000 Als Workers UE Directive 2004/40/EC [Electric Field] 10 % indici calcolati con picco ponderato
Zona 0 è la zona all’interno della quale i livelli di esposizione sono sicuramente inferiori o uguali ai livelli di riferimento della popolazione (conformità alla raccomandazione 1999/519/CE) Distanze maggiori di 0.9 m
Zona 1 è la zona all’interno della quale i livelli di esposizione superano i livelli di riferimento della popolazione ma sono inferiori o uguali ai valori di azione per i lavoratori stabiliti dalla Direttiva 2004/40/CE Distanze comprese tra 0.6 m e 0.9 m
Zona 2 è la zona nella quale i livelli di esposizione superano i livelli di azione stabiliti dalla Direttiva 2004/40/CE Distanze inferiori a 0.6 m

Note
Condizioni misura direzione assiale

Reference person:

AUSL 7 Siena - Laboratorio Agenti Fisici Andrea Bogi ([email protected])
SIZE AND SHAPE OF THE APPLICATOR [m]
X 0.1 SHAPE

Cylindrical
Y 0
Z 0.02
Measuring point Reference Distance Perpendicular [cm] Distance Parallel [cm] Height from ground [cm] Frequency dominant [Hz] FIGURE MEASURED ENTITY MEASURED VALUE NOTES
P1
applicatore
30 0 150 - Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 30000 % direzione perpendicolare alla spira
P2
applicatore
0 50 150 - Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 2900 % direzione parallela alla spira
P3
applicatore
50 0 150 - Low Als Workers UE Directive 2013/35/EC [Magnetic Field] 290 % direzione perpendicolare alla spira
P4
applicatore
0 50 150 - Low Als Workers UE Directive 2013/35/EC [Magnetic Field] 120 % direzione parallela alla spira
P5
applicatore
50 0 150 - High Als Workers UE Directive 2013/35/EC [Electric Field] 270 % direzione perpendicolare alla spira
P6
applicatore
0 50 150 - High Als Workers UE Directive 2013/35/EC [Electric Field] 110 % direzione parallela alla spira
Zona 0 è la zona all’interno della quale i livelli di esposizione sono sicuramente inferiori o uguali ai livelli di riferimento della popolazione (conformità alla raccomandazione 1999/519/CE) Distanze maggiori di 1.9 m
Zona 1 è la zona all’interno della quale i livelli di esposizione superano i livelli di riferimento della popolazione ma sono inferiori o uguali ai valori di AZIONE INFERIORI per i lavoratori stabiliti dal Decreto Legislativo n. 81/2008 Distanze comprese tra 0.7 m e 1.9 m
Zona 2 è la zona nella quale i livelli di esposizione superano i livelli di AZIONE INFERIORI per i lavoratori stabiliti dal Decreto Legislativo n. 81/2008 Distanze inferiori a 0.7 m

Note
logo regione toscana
Regione Toscana
Diritti Valori Innovazione Sostenibilità
collegamento sito usl 7 siena Servizio Sanitario