Machine technical sheet


Immagine rappresentane il macchinario
Brand: Elettronica Pagani
Model: RX 250
Type: Electromedical equipment: radar-therapy
Power supply: Electrical (220V-380V)
Reference standard: CEI EN 60601-2-3
Frequency of work: 2 - 300 GHz
Type of emission: Continua
Type of radiant element: Non disponibile
Category according to 12198-1-12009 Standard: Non disponibile
Required protection measures:
  1. Delimit Zone 1: Exposure values above the reference levels for general public
  2. Delimit Zone 2: Exposure values above the action levels for workers

Measurements in the field (Click to view the measurements in the field)

Reference person:

AUSL 7 Siena - Laboratorio Agenti Fisici Andrea Bogi ([email protected])
SIZE AND SHAPE OF THE APPLICATOR [m]
X 17 SHAPE

Cylindrical
Y 17
Z 16
Measuring point Reference Distance Perpendicular [cm] Distance Parallel [cm] Height from ground [cm] Frequency dominant [Hz] FIGURE MEASURED ENTITY MEASURED VALUE NOTES
P1
applicatore
30 20 - - Electric Field 140 V/m
P2
applicatore
60 20 - - Electric Field 60 V/m
P3
quadro comandi
35 0 - - Electric Field 60 V/m
Zona 0 è la zona all’interno della quale i livelli di esposizione sono sicuramente inferiori o uguali ai livelli di riferimento della popolazione (conformità alla raccomandazione 1999/519/CE) Distanze maggiori di 0.6 m
Zona 1 è la zona all’interno della quale i livelli di esposizione superano i livelli di riferimento della popolazione ma sono inferiori o uguali ai valori di AZIONE INFERIORI per i lavoratori stabiliti dal Decreto Legislativo n. 81/2008 Distanze comprese tra 0.3 m e 0.6 m
Zona 2 è la zona nella quale i livelli di esposizione superano i livelli di AZIONE INFERIORI per i lavoratori stabiliti dal Decreto Legislativo n. 81/2008 Distanze inferiori a 0.3 m

Note
frequenza di emissione 2450MHz