Machine technical sheet


Immagine rappresentane il macchinario
Brand: EME Srl - Electronic Medical Equipment
Model: Magnetomed 8200
Type: Electromedical equipment: magnetotherapy
Power: 0.41 kW
Power supply: Electrical (220V-380V)
Reference standard: CEI EN 60601-1
Frequency of work: 0 - 1000 Hz
Type of emission: Pulsata
Type of radiant element: Induttivo
Category according to 12198-1-12009 Standard: Non disponibile
Required protection measures:
  1. Delimit Zone 1: Exposure values above the reference levels for general public
  2. Operate by keeping the cranium and the trunk at a distance greater than the hazard distance
  3. The access to the Restricted areas shall be permitted only to those individuals with proven absence of contraindications to EMF exposure
  4. The access to the Restricted area is forbidden to individuals wearing pacemakers and/or implanted electronic devices

Measurements in the field (Click to view the measurements in the field)

Reference person:

AUSL Toscana Sud Est-LSP-Lab. Agenti Fisici Iole Pinto - Nicola Stacchini - Andrea Bogi ([email protected])
SIZE AND SHAPE OF THE APPLICATOR [m]
X 0.18 SHAPE

Cylindrical
Y 0.335
Z 0.335
Measuring point Reference Distance Perpendicular [cm] Distance Parallel [cm] Height from ground [cm] Frequency dominant [Hz] FIGURE MEASURED ENTITY MEASURED VALUE NOTES
P1
applicatore
46 0 - 50 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P3
applicatore
0 48 - 50 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P5
applicatore
28 0 - 50 Als Workers UE Directive 2004/40/EC [Magnetic Field] 100 %
P6
applicatore
0 34 - 50 Als Workers UE Directive 2004/40/EC [Magnetic Field] 100 %
Zona 0 è la zona all’interno della quale i livelli di esposizione sono sicuramente inferiori o uguali ai livelli di riferimento della popolazione (conformità alla raccomandazione 1999/519/CE) Distanze maggiori di 0.28 m
Zona 1 è la zona all’interno della quale i livelli di esposizione superano i livelli di riferimento della popolazione ma sono inferiori o uguali ai valori di AZIONE INFERIORI per i lavoratori stabiliti dal Decreto Legislativo n. 81/2008 Distanze comprese tra 0.1 m e 0.28 m
Zona 2 è la zona nella quale i livelli di esposizione superano i livelli di AZIONE INFERIORI per i lavoratori stabiliti dal Decreto Legislativo n. 81/2008 Distanze inferiori a 0.1 m

Note
Diametro spira 67 cm - Distanza parallela= distanza dal centro bobina lungo l'asse della stessa - Distanza Perpendicolare= distanza dal centro della bobina lungo il raggio. Zona 2 riferita bordo bobina zona 1 riferita bordo bobina

Reference person:

AUSL Toscana Sud Est-LSP-Lab. Agenti Fisici Iole Pinto - Nicola Stacchini - Andrea Bogi ([email protected])
SIZE AND SHAPE OF THE APPLICATOR [m]
X 0.36 SHAPE

Cylindrical
Y 0.67
Z 0.05
Measuring point Reference Distance Perpendicular [cm] Distance Parallel [cm] Height from ground [cm] Frequency dominant [Hz] FIGURE MEASURED ENTITY MEASURED VALUE NOTES
P1
applicatore
28 0 60 50 Low Als Workers UE Directive 2013/35/EC [Magnetic Field] 40 %
P2
applicatore
33 0 60 50 Low Als Workers UE Directive 2013/35/EC [Magnetic Field] 30 %
P3
applicatore
48 0 60 50 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P4
applicatore
0 33 60 50 Low Als Workers UE Directive 2013/35/EC [Magnetic Field] 40 %
P5
applicatore
0 44 60 50 Low Als Workers UE Directive 2013/35/EC [Magnetic Field] 20 %
P5
applicatore
0 44 60 50 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
Zona 0 è la zona all’interno della quale i livelli di esposizione sono sicuramente inferiori o uguali ai livelli di riferimento della popolazione (conformità alla raccomandazione 1999/519/CE) Distanze maggiori di 0.3 m
Zona 1 è la zona all’interno della quale i livelli di esposizione superano i livelli di riferimento della popolazione ma sono inferiori o uguali ai valori di AZIONE INFERIORI per i lavoratori stabiliti dal Decreto Legislativo n. 81/2008 Distanze comprese tra 0.1 m e 0.3 m
Zona 2 è la zona nella quale i livelli di esposizione superano i livelli di AZIONE INFERIORI per i lavoratori stabiliti dal Decreto Legislativo n. 81/2008 Distanze inferiori a 0.1 m

Note
logo regione toscana
Regione Toscana
Diritti Valori Innovazione Sostenibilità
collegamento sito usl 7 siena Servizio Sanitario
della
Toscana