Machine technical sheet


Immagine rappresentane il macchinario
Brand: Pozzi Elettromedicali Snc - Zola Predosa (BO)
Model: Magnetoterapix P1
Type: Electromedical equipment: magnetotherapy
Power: 0.035 kW
Power supply: Electrical (220V-380V)
Reference standard: CEI EN 60601-1
Frequency of work: 0 - 1000 Hz
Type of emission: Pulsata
Type of radiant element: Induttivo
Category according to 12198-1-12009 Standard: Non disponibile
Required protection measures:
  1. Delimit Zone 1: Exposure values above the reference levels for general public
  2. EMF emissions under the action levels for workers
  3. The access to the Restricted areas shall be permitted only to those individuals with proven absence of contraindications to EMF exposure
  4. The access to the Restricted area is forbidden to individuals wearing pacemakers and/or implanted electronic devices

Measurements in the field (Click to view the measurements in the field)

Reference person:

AUSL Toscana Sud Est-LSP-Lab. Agenti Fisici Iole Pinto - Nicola Stacchini - Andrea Bogi ([email protected])
SIZE AND SHAPE OF THE APPLICATOR [m]
X 0.22 SHAPE

Cubical
Y 0.26
Z 0.07
Measuring point Reference Distance Perpendicular [cm] Distance Parallel [cm] Height from ground [cm] Frequency dominant [Hz] FIGURE MEASURED ENTITY MEASURED VALUE NOTES
P1
applicatore
8 0 - 50 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P1
applicatore
8 0 - 50 Magnetic Field 12 microT
P3
applicatore
0 14 - 50 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P7
Non disponibile
0 0 - 50 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P8
applicatore
0 0 - 50 Als Workers UE Directive 2004/40/EC [Magnetic Field] 88 %
P9
Non disponibile
0 11 - 50 Als Workers UE Directive 2004/40/EC [Magnetic Field] 88 %
Zona 0 è la zona all’interno della quale i livelli di esposizione sono sicuramente inferiori o uguali ai livelli di riferimento della popolazione (conformità alla raccomandazione 1999/519/CE) Distanze maggiori di 0.08 m
Zona 1 è la zona all’interno della quale i livelli di esposizione superano i livelli di riferimento della popolazione ma sono inferiori o uguali ai valori di azione per i lavoratori stabiliti dalla Direttiva 2004/40/CE Distanze comprese tra 0.01 m e 0.08 m
Zona 2 è la zona nella quale i livelli di esposizione superano i livelli di azione stabiliti dalla Direttiva 2004/40/CE Distanze inferiori a 0.01 m

Note
Distanza parallela= distanza dal centro bobina lungo l'asse della stessa - Distanza Perpendicolare= distanza dal centro della bobina lungo il raggio. Zona 1-2 riferita bordo bobina zona 1 riferita bordo bobina

Reference person:

AUSL Toscana Sud Est-LSP-Lab. Agenti Fisici Iole Pinto - Nicola Stacchini - Andrea Bogi ([email protected])
SIZE AND SHAPE OF THE APPLICATOR [m]
X 0.16 SHAPE

Parallelepipedal
Y 0.2
Z 0.05
Measuring point Reference Distance Perpendicular [cm] Distance Parallel [cm] Height from ground [cm] Frequency dominant [Hz] FIGURE MEASURED ENTITY MEASURED VALUE NOTES
P1
lato destro apparato
10 0 0 40 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P2
lato sinistro apparato
10 0 0 40 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P3
parete laterale
0 5 0 40 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P4
parete laterale
0 5 0 40 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P5
parete laterale
0 0 5 40 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P6
parete laterale
4 0 0 40 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P7
parete laterale
0 2 0 40 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
P8
parete laterale
0 0 2 40 Reference level General Public council racommendation 1999/519/EC [Magnetic Field] 100 %
Zona 0 è la zona all’interno della quale i livelli di esposizione sono sicuramente inferiori o uguali ai livelli di riferimento della popolazione (conformità alla raccomandazione 1999/519/CE) Distanze maggiori di 0.1 m
Zona 1 è la zona all’interno della quale i livelli di esposizione superano i livelli di riferimento della popolazione ma sono inferiori o uguali ai valori di azione per i lavoratori stabiliti dalla Direttiva 2004/40/CE Distanze comprese tra 0.05 m e 0.1 m
Zona 2 è la zona nella quale i livelli di esposizione superano i livelli di azione stabiliti dalla Direttiva 2004/40/CE Distanze inferiori a 0.05 m

Note
Condizioni di misura: programma fratture osteoporosi Frequenza 40 Hz