Machine technical sheet


Immagine rappresentane il macchinario
Brand: VILBER LOURMAT Deutschland GmbH
Model: VL-4C
Type: Transilluminator
Power supply: Electrical (220V-380V)
Reference standard: NON DISPONIBILE
Classification according to standard IEC-EN 62471 / IEC-EN 60825-1: Non disponibile
Category according to standard EN12198-1 Non disponibile

SOURCE TECHNICAL SHEET
Type: lampada a scarica di GAS
Emission: Continua
N. of lighting bodies: 2
Brand: NON DISPONIBILE
Model: NON DISPONIBILE
Classification according to standard IEC-EN 62471 / IEC-EN 60825-1: Non disponibile
Reference standard: NON DISPONIBILE

RELEVANT EMISSION TYPES  
                                       
            780 ÷ 3000 (IRA e IRB)    
            780 ÷ 1400 (IRA)                    
        380 ÷ 1400
(Visible and IRA)
                   
      315 ÷ 400
(UVA)
                             
    300 ÷ 700
(Blue light)
                             
  180 ÷ 400
(UVA, UVB e UVC)
                             
⇐ BAND ⇒    
Main protection measures to implement for safety purposes:
  1. Operators should protect their eyes and skin

Measurements in the field (Click to view the measurements in the field)

Reference person:
AUSL 7 Siena - Laboratorio Agenti Fisici Andrea Bogi ([email protected])

Rif.D.lgs 81/08 SPECTRAL RANGE TARGET ORGANS RESULT Val.
a ES (Ultraviolet) Eyes and skin Maggiore del VLE 0.02 W m-2
b EUVA (Ultraviolet A) Eyes Irrilevante 0.003 W m-2
c,d LB (Blue light, extended source) Eyes Irrilevante
e,f EB (Blue light, small source) Eyes Non applicabile
  LAFA (Blue light, extended source, aphakic individuals) Eyes Irrilevante
  EAFA (Blue light, small source,aphakic individuals) Eyes Non applicabile
g,h,i L R (Visible and Infrared A) Eyes Irrilevante
j,k,l L R (Infrared A) Eyes Irrilevante
m,n EIR (Infrared A + Infrared B) Eyes Irrilevante
o Eskin (Visible + Infrared A + Infrared B) Skin Irrilevante
SENSOR-SOURCE DISTANCE
17
cm
MAXIMUM EXPOSURE TIME
1300
seconds
SAFETY RADIAL DISTANCE
100
cm
Note: misura riflessa sui campioni, lunghezza d�onda selezionata 254 nm
Reference person:
AUSL 7 Siena - Laboratorio Agenti Fisici Andrea Bogi ([email protected])

Rif.D.lgs 81/08 SPECTRAL RANGE TARGET ORGANS RESULT Val.
a ES (Ultraviolet) Eyes and skin Irrilevante 0 W m-2
b EUVA (Ultraviolet A) Eyes da 20% � 50% del limite 0.1 W m-2
c,d LB (Blue light, extended source) Eyes Irrilevante
e,f EB (Blue light, small source) Eyes Non applicabile
  LAFA (Blue light, extended source, aphakic individuals) Eyes Irrilevante
  EAFA (Blue light, small source,aphakic individuals) Eyes Non applicabile
g,h,i L R (Visible and Infrared A) Eyes Irrilevante
j,k,l L R (Infrared A) Eyes Irrilevante
m,n EIR (Infrared A + Infrared B) Eyes Irrilevante
o Eskin (Visible + Infrared A + Infrared B) Skin Irrilevante
SENSOR-SOURCE DISTANCE
18
cm
SAFETY RADIAL DISTANCE
20
cm
Note: misura riflessa sui campioni, lunghezza d�onda selezionata 365 nm
logo regione toscana
Regione Toscana
Diritti Valori Innovazione Sostenibilità
collegamento sito usl 7 siena Servizio Sanitario
della
Toscana