Machine technical sheet


Immagine rappresentane il macchinario
Brand: Tutti
Model: tutti
Type: Germicidal lamp LED UVC
Power supply: Battery 1V-7,2V
Reference standard: CEI EN 62471
Classification according to standard IEC-EN 62471 / IEC-EN 60825-1: 3
Category according to standard EN12198-1 Non disponibile

SOURCE TECHNICAL SHEET
Type: LED
Emission: Continua
N. of lighting bodies: 3
Brand: lunghezza d'onda centrale 278nm
Model: pot radiante 15-18 mW
Classification according to standard IEC-EN 62471 / IEC-EN 60825-1: Non disponibile
Reference standard: CEI 62-39

RELEVANT EMISSION TYPES  
                                       
            780 ÷ 3000 (IRA e IRB)    
            780 ÷ 1400 (IRA)                    
        380 ÷ 1400
(Visible and IRA)
                   
      315 ÷ 400
(UVA)
                             
    300 ÷ 700
(Blue light)
                             
  180 ÷ 400
(UVA, UVB e UVC)
                             
⇐ BAND ⇒    
Main protection measures to implement for safety purposes:
  1. Limits exceeded even for reflected radiation.
  2. It is mandatory to train the operators: the exposure is relevant if the operators are not informed about the appropriate operation modes

Measurements in the field (Click to view the measurements in the field)

Reference person:
AUSL 7 Siena - Laboratorio Agenti Fisici Andrea Bogi ([email protected])

Rif.D.lgs 81/08 SPECTRAL RANGE TARGET ORGANS RESULT Val.
a ES (Ultraviolet) Eyes and skin Maggiore del VLE 0 W m-2
b EUVA (Ultraviolet A) Eyes Maggiore del VLE 1 W m-2
c,d LB (Blue light, extended source) Eyes Non applicabile
e,f EB (Blue light, small source) Eyes Maggiore del VLE 0.3 W / m2
  LAFA (Blue light, extended source, aphakic individuals) Eyes Non applicabile
  EAFA (Blue light, small source,aphakic individuals) Eyes Maggiore del VLE 4 W m2
g,h,i L R (Visible and Infrared A) Eyes Irrilevante
j,k,l L R (Infrared A) Eyes Irrilevante
m,n EIR (Infrared A + Infrared B) Eyes Irrilevante
o Eskin (Visible + Infrared A + Infrared B) Skin Irrilevante
SENSOR-SOURCE DISTANCE
5
cm
MAXIMUM EXPOSURE TIME
25
seconds
Reference person:
AUSL 7 Siena - Laboratorio Agenti Fisici Andrea Bogi ([email protected])

Rif.D.lgs 81/08 SPECTRAL RANGE TARGET ORGANS RESULT Val.
a ES (Ultraviolet) Eyes and skin Maggiore del VLE 0.05 W m-2
b EUVA (Ultraviolet A) Eyes da 20% ÷ 50% del limite 0.1 W m-2
c,d LB (Blue light, extended source) Eyes Non applicabile
e,f EB (Blue light, small source) Eyes Maggiore del VLE 0.02 W / m2
  LAFA (Blue light, extended source, aphakic individuals) Eyes Non applicabile
  EAFA (Blue light, small source,aphakic individuals) Eyes Maggiore del VLE 0.2 W m2
g,h,i L R (Visible and Infrared A) Eyes Irrilevante
j,k,l L R (Infrared A) Eyes Irrilevante
m,n EIR (Infrared A + Infrared B) Eyes Irrilevante
o Eskin (Visible + Infrared A + Infrared B) Skin Irrilevante
SENSOR-SOURCE DISTANCE
20
cm
MAXIMUM EXPOSURE TIME
430
seconds
SAFETY RADIAL DISTANCE
150
cm
Reference person:
AUSL 7 Siena - Laboratorio Agenti Fisici Andrea Bogi ([email protected])

Rif.D.lgs 81/08 SPECTRAL RANGE TARGET ORGANS RESULT Val.
a ES (Ultraviolet) Eyes and skin Maggiore del VLE 0.07 W m-2
b EUVA (Ultraviolet A) Eyes da 20% ÷ 50% del limite 0.1 W m-2
c,d LB (Blue light, extended source) Eyes Non applicabile
e,f EB (Blue light, small source) Eyes Maggiore del VLE 0 W / m2
  LAFA (Blue light, extended source, aphakic individuals) Eyes Non applicabile
  EAFA (Blue light, small source,aphakic individuals) Eyes Maggiore del VLE 0 W m2
g,h,i L R (Visible and Infrared A) Eyes Irrilevante
j,k,l L R (Infrared A) Eyes Irrilevante
m,n EIR (Infrared A + Infrared B) Eyes Irrilevante
o Eskin (Visible + Infrared A + Infrared B) Skin Irrilevante
SENSOR-SOURCE DISTANCE
14
cm
MAXIMUM EXPOSURE TIME
370
seconds
Reference person:
AUSL 7 Siena - Laboratorio Agenti Fisici Andrea Bogi ([email protected])

Rif.D.lgs 81/08 SPECTRAL RANGE TARGET ORGANS RESULT Val.
a ES (Ultraviolet) Eyes and skin Maggiore del VLE 0.004 W m-2
b EUVA (Ultraviolet A) Eyes Irrilevante
c,d LB (Blue light, extended source) Eyes Irrilevante
e,f EB (Blue light, small source) Eyes Non applicabile
  LAFA (Blue light, extended source, aphakic individuals) Eyes Irrilevante
  EAFA (Blue light, small source,aphakic individuals) Eyes Non applicabile
g,h,i L R (Visible and Infrared A) Eyes Irrilevante
j,k,l L R (Infrared A) Eyes Irrilevante
m,n EIR (Infrared A + Infrared B) Eyes Irrilevante
o Eskin (Visible + Infrared A + Infrared B) Skin Irrilevante
SENSOR-SOURCE DISTANCE
9
cm
MAXIMUM EXPOSURE TIME
7400
seconds
logo regione toscana
Regione Toscana
Diritti Valori Innovazione Sostenibilità
collegamento sito usl 7 siena Servizio Sanitario
della
Toscana